Influence of pore structure of silica packing on hplc column characteristics
I. Novák, B. Buszewski, J. Garaj, and D. Berek
Polymer Institute, Centre for Chemical Research,
Slovak Academy of Sciences, CS-84236 Bratislava
Abstract: The influence of the inner structure of bare silica gel sorbents — especially of their pore volume — on the characteristics of the resulting HPLC
columns was studied. The sorbents of various pore characteristics were
tested and packed into the columns at optimized column packing conditions.
The results show that an increase in the sorbent pore volume leads to a
slight increase of the column permeability and to a rather pronounced rise
of the sample capacity expressed in terms of volume and concentration of
a sample. Highly efficient and fairly stable columns can be obtained with
strictly spherical silica gels even if their specific pore volume reaches the
value of 2 cm3 g-1.
Full paper in Portable Document Format: 441a31.pdf
Chemical Papers 44 (1) 31–43 (1990)
|