Amperometrically indicated "pseudotitrations". 4. Complexometric determination of Cr(III) compounds in low concentrations using deconvolution voltammetry and differential pulse polarography
D. Bustin and M. Rievaj
Department of Analytical Chemistry, Slovak Technical University,
CS-81237 Bratislava
Abstract: The Cr(III) compounds can be determined complexometrically by the
direct "pseudotitration" in spite of the fact that the complexation reaction
with EDTA is very slow and its range in the bulk phase of titrated solution
is negligible. In the vicinity of amperometrical indicating electrode, however,
the titration reaction takes place sufficiently rapidly due to the effective
redox catalysis. Application of deconvolution voltammetry and differential
pulse polarography enables to determine Cr(III) in the minimal concentration 5 x 10-6 mol dm-3 Reasons are given which make it impossible to
achieve mpre favourable (lower) limit of determination.